| Input |
|
| Pressure
Range |
4 bar to 690
bar |
| Proof
Pressure |
2 x Full
Scale (FS) for Stainless Steel Units |
| |
1.5 x FS for
Inconel Units |
| Burst
Pressure |
>35 x Fs
<= 10 bar ranges |
| |
>15 x FS
<= 100 bar ranges |
| |
>8 FS
<= 690 bar ranges |
| Fatigue
Life |
3 million FS
cycles |
| Performance |
|
| Output |
4- 20 mA (2
wire) |
| Supply
Voltage (Vs) |
8 to 40 Vdc |
| Supply
Voltage Sensitivity |
0.005% of
max. span/ Volt |
| Long Term
Drift |
0.1% of max
span per year non- cumulative |
| Accuracy |
0.1 % FS
typical |
| Thermal
Error (typical) |
0.8% of max
span for performance code E |
| Compensated
Temperatures |
-25° to 75°
C (- 13° to 167° F) |
| Operating
Temperatures |
-25° to 85°
C (- 13° to 185° F) |
| Zero
Tolerance |
0.1% FS,
typical |
| Span
Tolerance |
0.1% FS,
typical |
| Zero
Adjustment |
+/- 10% (100%
at factory) by potentiometer |
| Span
Adjustment |
25% to 125 %
of span by potentiometer |
| Max. Loop
Resistance |
(Vs- 8) x 50
ohms |
|
| Mechanical
Configuration |
|
| Pressure
Port |
see ordering
chart |
| Wetted
Parts |
17- 4 PH ss
(optional Inconel) |
| |
[17- 4 PH and
15- 7 Mo Stainless Steel <= 1.6 bar (30 Psi)] |
| Electrical
Connection |
M20 thread
giving access to terminal blocks, optional
flameproof cable assembly |
| Enclosure |
321 ss, 17- 4
PH ss |
| |
IP50 when
used with approved cable assembly |
| Vibration |
35g peak
sinusoidal, 5 to 2000 Hz |
| Acceleration |
100g steady
acceleration in any direction 0.05% FS/ g for 1
bar (15 psi) range decreasing logarthmicaly to
0.0001% FS/ g for 690 bar (10000 psi) range. |
| Shock |
Withstands
free fall to IEC 68- 2- 32 procedure 1 |
| Approvals |
CE,
Flameproof EEx d IIC T4 per CENELEC |
| Weight |
approx. 1.5Kg
|
|